Original research Open Access Logo

Development and Validation of a Sensitive RP-HPLC Method for Quantifying Oxaliplatin in ZIF-8 as a Drug Delivery System

Quyen Toan Pham 1, 2, 3
Xuan Hiep Nguyen 1, 2
Nhat Hoang Pham 1, 2
Linh Ho Thuy Nguyen 1, 2
Tri Minh Le 1, 2, 3
Tan Le Hoang Doan 1, 4, *
  1. Vietnam National University Ho Chi Minh City, Ho Chi Minh city, Viet Nam
  2. University of Health Sciences, Vietnam National University Ho Chi Minh City, Ho Chi Minh City, Viet Nam
  3. Faculty of Pharmacy, University of Medicine and Pharmacy at Ho Chi Minh City, Ho Chi Minh City, Vietnam
  4. Center for Innovative Materials and Architectures, Ho Chi Minh City, Viet Nam
Correspondence to: Tan Le Hoang Doan, Vietnam National University Ho Chi Minh City, Ho Chi Minh city, Viet Nam; Center for Innovative Materials and Architectures, Ho Chi Minh City, Viet Nam. Email: [email protected].
Volume & Issue: Vol. 6 No. 2 (2025) | Page No.: 795-804 | DOI: 10.32508/stdjhs.v6i2.602
Published: 2025-12-31

Online metrics


Statistics from the website

  • Abstract Views: 0
  • Galley Views: 0

Statistics from Dimensions

Copyright The Author(s) 2018. This article is published with open access by Vietnam National University, Ho Chi Minh city, Vietnam. This article is distributed under the terms of the Creative Commons Attribution License (CC-BY 4.0) which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited. 

Abstract

Introduction: Oxaliplatin is widely utilized in cancer therapy but is often limited by significant side effects. Zeolitic Imidazolate Framework-8 (ZIF-8), a type of metal-organic framework nanoparticle, has been investigated as a delivery system for oxaliplatin, offering advantages such as targeted cancer cell delivery and controlled drug release. This study aimed to develop and validate a sensitive, straightforward, and cost-effective reverse-phase high-performance liquid chromatography (RP-HPLC) method for quantifying oxaliplatin within ZIF-8.

Methods: Various chromatographic conditions were optimized to accurately quantify oxaliplatin-loaded ZIF-8. The final HPLC method was validated in accordance with International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use (ICH) guidelines.

Results: Chromatographic separation was achieved on an RP-C18 column (250 x 4.6 mm, 5 µm) using isocratic elution with a methanol-water mobile phase (1:1 ratio) at a flow rate of 0.8 mL/min. Detection was performed with a PDA detector set at 230 nm. The method demonstrated suitability through validation of specificity, linearity, accuracy, precision, range, and robustness. Linearity was confirmed within a concentration range of 0.1-1.5 mg/mL, with a correlation coefficient (R²) of 0.9994. Accuracy ranged from 98.79% to 99.71%, with a mean accuracy of 99.16%, and the relative standard deviation was 0.48%.

Conclusion: The validated RP-HPLC method offers a reliable and robust means for assessing the loading capacity and release performance of oxaliplatin in ZIF-8, serving as an essential analytical tool for evaluating this drug delivery system. This approach supports precise formulation assessment, aiding in the accelerated development of future applications in drug delivery research.